| Rheometer Installation Requirements
2. 110V 15A electrical circuit with a minimum of 6
receptacles (or 2 receptacles and a power strip). (*Note: A dedicated, isolated circuit and circuit breaker with variation in line voltage of less that 1% is required in order to avoid inadequate power situations. 3. Five (5) feet of usable bench space with access below bench top for temperature control circulator (if ordered) and drive motor cooling system. 4. Pentium computer or equivalent, 256MB ram (512MB ram is required for the Fast Oscillation test mode), 3.5" drive, SVGA monitor, mouse, and Windows 2000, or XP operating systems. 5. Water and Oil filters for Air line Notes: Filter (B) is downstream of filter (A). ATS RheoSystems requires a 3/8 NPT female fitting (standard on filter B) for connection to the Rheometer. Air filters must be installed prior to arrival date of instrument installation
7. For STRESSTECH/DSR QC, 1 open com port (9 pin sub D) is required.
|
|||||||||||||||
![]() ![]() |
||
|